Preset the right Scheduled Station AE Title on local ultrasound system according to
the server settings.
The ultrasound system is defaulted only to inquire the patient information that the
Modality is US, and the patient needs the server to provide US attribute. Change the
query conditions (e.g. date or patient name) to make the server find the record
accordingly. Based on DICOM standard, Patient ID cannot support fuzzy inquiry, only
can support the whole word match. Please confirm the entered Patient ID is right or
not.
3.5
Common failure: Printing succeeded
but the film is blank
Potential Reasons:
The printer doesn’t support color printing which is performed by the ultrasound
system;
The film or paper of the printer doesn’t support color printing.
Troubleshooting:
Check the printer conformance statement if the printer can support color printing; if
not, modify the “Settings” in DICOM Service -> Print to B/W type (MONOCHROME2);
Generally, the "Medium Type” is set to “Blue File” for B/W printing while “Paper” or
“Clear Film” for color printing
3.6
Common Failure: Verification of
connecting DICOM printer is succeed
but printing is failed
Potential Reasons:
Wrong AE_TITLE settings on local device and the server.
The parameters set in DICOM Service -> Print are not supported
Troubleshooting:
Set the correct local AE_TITLE and server AE_TITLE on local device according to the
DICOM Conformance Statement or settings of the printer.
Refer to the printer conformance statement for the printing parameters (Format, Film
Size) value that supported. Check film size and type, and then correctly set Film Size,
Format, Medium Type etc. in print service preset in ultrasound system.
3.7
Common Failure: Printed image is too
small
Potential Reasons:
The Magnification Type of local ultrasound system is set to NONE while this
parameter on printer is set to same with the local ultrasound system.