mindray
DC7 Diagnostic Ultrasound System
12
(
Sec
)
Timeout
(
Sec
)
5-60s, in increments of 5s, and 15 is default. Timeout is the overtime
of connecting Association.
Scheduled
Station
AE
Title
To set the AE Title of the scheduled station is optional. If the
Scheduled Station of the Worklist server is the ultrasound system,
then AE Title here, AE Title of the scheduled station should be the
consistent with the one set in the ultrasound system.
Add
Click it to add the Worklist service to the service list.
Cancel
Click to cancel parameter preset.
Update
Select an item in the service list, change the parameters in the above
area, and click [Update] to update the item in the service list.
Delete
Click to delete the selected service in the service list.
Verify
Click to verify if the two DICOM application entities are normally
connected.
Default
Select an item in the service list, click [Default] and you can see “Y”
in the Default column.
Exit
Click to exit the setup screen.
Tip:
In terms of “Scheduled Station AE Title”, if you set this item in the Worklist server, then
“Scheduled Station AE Title” configured here should be consistent with the one set in the
server. For example, when you set “Scheduled Station AE Title” in the Worklist server as
“Mindray” (this means you assigned the scheduled exam to “Mindray”), then both
“Scheduled Station AE Title” and “AE Title” in the ultrasound system configuration should
be “Mindray”.