5-16 Function Checking and Testing
NOTE:
1.
The measurement caliper should be positioned at the upper edge of
the target, not the middle nor the lower edge.
2.
The scanning plane should be vertical to the target line, that means
the scanning plane is parallel with the cross-section of the phantom
As shown in figure below.
Transverse geometric positioning accuracy
Test Step:
1. Cover the scan surface of the phantom with water or couple gel, gently contact the
probe with the scan surface
2. Adjust the depth, making the transverse targets to be displayed in the image.
3. Adjust the focus point to be posited beside the transverse targets (the standard is
not clear)
4. Adjust parameters like gain, TGC, making each transverse targets to be clearly
displayed.
5. Record the distance by 20mm each segment on the transverse targets line by using
the measurement caliper
6. Select the value with the greatest error (to 20mm), calculate the accuracy by using
the formula below
NOTE:
1.
When using a linear probe, record the transverse distance by
segment.
2.
When using a convex probe, all transverse targets should be
displayed integrally in an image.
Summary of Contents for DC-25
Page 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Page 2: ......
Page 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Page 8: ......
Page 12: ......
Page 70: ......
Page 88: ...5 18 Function Checking and Testing ...
Page 94: ...6 6 Software installation and Maintenance ...
Page 168: ......
Page 178: ......
Page 190: ......
Page 206: ......
Page 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Page 212: ...D 2 Boot Screen Doppler Boot Screen ...
Page 213: ...P N 046 009019 00 11 0 ...