Troubleshooting of Regular Malfunctions 11-7
3
Only one speaker is mute.
Speaker errors
Replace the speaker.
4
USB device cannot be
recognized while
connecting to the USB
port
Control panel works well;
Main board malfunction
Replace the main
board.
5
Network disconnected or
abnormal
CPU module or IO interface
board malfunction if there is
no network or setting failure.
Confirm by replacing
CPU module and IO
interface board
11.6 Control Panel
11.6.1 Module or Board Related
No.
Descriptions
Remarks
1
Silicon key
2
Control panel
3
Dual encoder module
4
Single encoder module
5
Trackball module
6
TGC module
11.6.2 Key Points Supporting Troubleshooting
No.
Key Points Supporting Troubleshooting
Remarks
1
Backlight of control panel
To confirm if the control panel is
powered on normally;
2
Key sound of the control panel.
To confirm if the buzzer works normally
3
Response to function keys on control
panel
4
Response to trackball operation
5
Response to the sliding potentiometer
To
confirm
if
a
single
sliding
potentiometer or the control panel
PCBA is damaged.
6
Response to each encoder
To confirm if a single encoder or the
control panel PCBA is damaged.
Summary of Contents for DC-25
Page 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Page 2: ......
Page 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Page 8: ......
Page 12: ......
Page 70: ......
Page 88: ...5 18 Function Checking and Testing ...
Page 94: ...6 6 Software installation and Maintenance ...
Page 168: ......
Page 178: ......
Page 190: ......
Page 206: ......
Page 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Page 212: ...D 2 Boot Screen Doppler Boot Screen ...
Page 213: ...P N 046 009019 00 11 0 ...