11-6 Troubleshooting of Regular Malfunctions
All probes cannot be recognized in all probe
sockets;
Probe board ID cannot be read correctly by the
system information;
Main board or probe board malfunction.
Replace the
probe board or
main board
assembly.
11.5 IO Interface System
11.5.1 Module or Board Related
No.
Descriptions
Remarks
1
Main board
2
CPU module
3
IO interface board
Failure rate: low
4
IO Rear board
Failure rate: low
11.5.2 Key Points Supporting Troubleshooting
No.
Key Points Supporting Troubleshooting
Remarks
1
Working condition of the control panel
Confirm if the USB main device is
functional since the control panel is
connected by internal USB cable
2
Working condition of each USB port
3
Working condition of Video, S-Video and VGA
video interfaces
4
Printing condition of video printer
Check if output of connected video
printer is normal
11.5.3 Troubleshooting
No.
Fault Description
Cause Analysis
Measure
1
LCD displays normally;
NO video printer output
when press <print> on
control panel;
Remote
printing
control
failure or video output failure
if there is no user-defined
key failure.
Replace the main
board.
2
Two speakers are mute.
Errors of the main board.
Replace the main
board.
Summary of Contents for DC-25
Page 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Page 2: ......
Page 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Page 8: ......
Page 12: ......
Page 70: ......
Page 88: ...5 18 Function Checking and Testing ...
Page 94: ...6 6 Software installation and Maintenance ...
Page 168: ......
Page 178: ......
Page 190: ......
Page 206: ......
Page 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Page 212: ...D 2 Boot Screen Doppler Boot Screen ...
Page 213: ...P N 046 009019 00 11 0 ...