System Diagnosis and Support
9-9
9.3.3
Temperature Alarming
9.3.4
Fan Alarming
the current value, and [LLL]-[HHH] represents the
lower and upper limits.
The voltage names respectively are:
+12V
、
-12V
、
-5V
、
2V5
、
1V8
、
1V8A
、
1V5
、
1V2.
Alarming tips
LOG record
Suggestion
Temperature
alarming, XX
Shut down
(XX means
inversion
timing,
starting 60S)
××× System Monitor: Temperature Alert!
[XXX], Current voltage: [VVV] degree, Limit
temperature: [LLL]-[HHH] degree
×××represents time, e.g. 2010-12-25
14:15:25
[XXX] represents voltage name, [VVV]
represents the current value, and
[LLL]-[HHH] represents the lower and upper
limits. The detailed temperature name is as
following:
ADC thermal sensor (Temperature
transducer of front AD), Multi FPGA thermal
sensor(Temperature transducer of MUL
FPGA)
Check the fan log
D:/DCN2PLus/Log/PeriLog
(if the fan stops
working)/check if the heat
radiating condition is
normal.
The log records are the same
The temperature name is: power thermal
sensor
as above
none
×××: System Monitor: Temperature Alert!
[CPU thermal sensor], Current voltage:
[VVV] degree, Limit temperature: [LLL]
degree
×××represents time, e.g. 2010-12-25
14:15:25
[VVV] represents the current value, and [LLL]
represents the first limit of CPU temperature.
The first temperature of
CPU
Suggestion as above
Temperature
alarming, XX
Shut down
(XX means
inversion
timing,
starting 60S)
The log records are the same
[VVV] represents the current value, and [LLL]
represents the second limit of CPU
temperature.
The second temperature of
CPU
Suggestion as above
Alarming tips
LOG record
Suggestion
Fans need
maintenance,
please contact
×××: System Monitor: Fan alert
!
[XXX], Current
speed: [VVV] degree, Limit speed: [LLL] rpm.
Replace the
fan/connect
the line
Summary of Contents for DC-25
Page 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Page 2: ......
Page 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Page 8: ......
Page 12: ......
Page 70: ......
Page 88: ...5 18 Function Checking and Testing ...
Page 94: ...6 6 Software installation and Maintenance ...
Page 168: ......
Page 178: ......
Page 190: ......
Page 206: ......
Page 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Page 212: ...D 2 Boot Screen Doppler Boot Screen ...
Page 213: ...P N 046 009019 00 11 0 ...