System Diagnosis and Support
9-11
9.3.6
PHV Related Alarming
Alarming tips
LOG record
Suggestion
Alarm!
High-voltage
transmission is
abnormal, and
images display
normally!
PHVX supply voltage error, PHVX
is XX(upper limit is XX)
Something is wrong with the
programmed voltage of power
module, which make either of the
both PHV circuits or multi-path
voltage 1.1 more than the limit of
voltage (100%AP
)
.
none
PHV protect, PHV1P is XX,
PHV1N is XX
PHV protection, and the voltage of
PHV is less than the lower limit
95V Power supply voltage error,
the voltage is XX ,normal region is
90V to 100V
Something is wrong with the
programmed voltage of power
module: positive/negative 95V of
power transmitting exceeds the
limited value
-95V Power supply voltage error,
the voltage is XX ,normal region is
-90V to -100V
Summary of Contents for DC-25
Page 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Page 2: ......
Page 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Page 8: ......
Page 12: ......
Page 70: ......
Page 88: ...5 18 Function Checking and Testing ...
Page 94: ...6 6 Software installation and Maintenance ...
Page 168: ......
Page 178: ......
Page 190: ......
Page 206: ......
Page 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Page 212: ...D 2 Boot Screen Doppler Boot Screen ...
Page 213: ...P N 046 009019 00 11 0 ...