5-4 Function Checking and Testing
5.3.2.4
Check ECG Module
Procedure
Standard
Confirm if the ECG module is
configured, then:
Set [ECG] as “On”.
ECG trace is displayed; the heart icon is
displayed at the upper right corner of the
screen.
The parameters [Speed], [ECG Gain] can be
adjusted.
ECG signal can be reviewed correctly.
5.3.2.5
Checking IO Interface
Procedure
Standard
Checking IO Interfaces
Video interface, USB port have
been verified, remains are:
VGA interface
Connect VGA display
Network port
USB port
The contents displayed on the VGA/LCD are
the same as those displayed on the
ultrasound system main monior, no character
and image loss, no color difference, no
fluttering and flicking.
Smooth communication in both network and
connection.
USB port data storage/accessing are normal.
5.4
Function Checks
NOTE:
A complete function inspection is described here, do the checking according to
the actual system configuration.
Summary of Contents for DC-25
Page 1: ...DC 32 DC 30 DC 28 DC 26 DC 25 Digital Ultrasound System Service Manual Revision 11 0 ...
Page 2: ......
Page 7: ...v Appendix C Requirements of Performance Indices C 1 Appendix D Boot Screen D 1 ...
Page 8: ......
Page 12: ......
Page 70: ......
Page 88: ...5 18 Function Checking and Testing ...
Page 94: ...6 6 Software installation and Maintenance ...
Page 168: ......
Page 178: ......
Page 190: ......
Page 206: ......
Page 211: ...Boot Screen D 1 Appendix D Boot Screen BIOS Boot Screen LINUX Boot Screen ...
Page 212: ...D 2 Boot Screen Doppler Boot Screen ...
Page 213: ...P N 046 009019 00 11 0 ...