A-3
Selected energy accuracy
Impedance
Energy
25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Accuracy
1
J
1 1 1 0.9 0.9 0.9 0.8 ±2J
2
J
2 2 2 1.9 1.8 1.7 1.6 ±2J
3
J
2.9 3 2.9 2.8 2.7 2.6 2.4 ±2J
4
J
3.9 4 3.9 3.7 3.6 3.4 3.2 ±2J
5
J
4.9 5 4.9 4.7 4.5 4.3 4.1 ±2J
6
J
5.8 6 5.8 5.6 5.3 5.1 4.9 ±2J
7
J
6.8 7 6.8 6.6 6.3 6
5.7 ±2J
8
J
7.8 8 7.8 7.4 7.1 6.8 6.5 ±2J
9
J
8.8 9 8.8 8.4 8
7.7 7.3 ±2J
10
J
9.7 10 9.7 9.3 8.9 8.5 8.1 ±2J
15
J
15 15 15 14 13 13 12 ±15%
20
J
20 20 20 19 18 17 16 ±15%
30
J
29 30 29 28 27 25 24 ±15%
50
J
49 50 49 47 45 43 41 ±15%
70
J
68 70 68 65 62 60 57 ±15%
100
J
97 100
97 93 89 85 81 ±15%
150
J
146 150 146 140 134 128 122 ±15%
170
J
166 170 166 159 151 145 138 ±15%
200
J
195 200 195 187 178 170 163 ±15%
300
J
292 300 292 280 267 255 244 ±15%
360
J
351 360 350 336 321 306 293 ±15%
Charge time (Note: at 20
C of ambient temperature)
Charge time
(Manual Defib)
From initiation of rhythm
analysis to charge done
From initial power
on to charge done
200J
360J
200J
360J
200J
360J
Rechargeabl
e battery
New and fully charged
<5 s
<8 s
<10 s
<10 s
<20 s
<20 s
New and fully charged after
15 times of 360J discharges
<6 s
<9 s
<10 s
<10 s
<20 s
<20 s
Disposable
battery
New
/
/
<10 s
<17 s
<20 s
<27 s
New after 15 times of 360J
discharges
/
/
<10 s
<17 s
<20 s
<27 s
Synchronized discharge delay
Local synchronized discharge delay
< 60ms (from the peak of R-wave)
Summary of Contents for BeneHeart D1
Page 1: ...BeneHeart D1 Automated External Defibrillator Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 42: ...6 6 FOR YOUR NOTES...
Page 50: ...7 8 FOR YOUR NOTES...
Page 58: ...9 6 FOR YOUR NOTES...
Page 86: ...B 4 FOR YOUR NOTES...
Page 88: ...C 2 FOR YOUR NOTES...
Page 92: ...D 4 FOR YOUR NOTES...
Page 96: ...E 4 FOR YOUR NOTES...
Page 100: ...F 4 FOR YOUR NOTES...
Page 102: ...G 2 FOR YOUR NOTES...
Page 104: ......
Page 105: ......
Page 106: ...P N 046 004673 00 3 0...