12-4
12.3.5
User Test
WARNING
Do not perform user test when a patient is connected to the equipment.
On the maintenance screen, you can press [
User Test
] to perform user test. You need to:
1.
Press the corresponding soft keys and
Shock
button as indicated on the screen.
2.
Select an option to respond to the equipment if you have heard the voice prompt.
Other items are tested hereafter automatically. If all items pass the test, the test result is “Test Passed”. If any failure is
detected, corresponding prompts and failure codes are displayed.
12.3.6
Electrical Safety Tests
For details about Electrical Safety Tests, refer to Appendix
E Electrical Safety Inspection
.