7-4
Pads Placement for Pediatric Patient
Place the pink pad in the center of the chest between the nipples, and the yellow one in the center of the back
(anterior-posterior), as shown below:
7.5
Changing ECG Settings
7.5.1
Selecting Lead Type
In Manual Defibrillation mode, select [
>>
] and then press [
Lead
] repeatedly to set lead type as per the adopted lead
type.
7.5.2
Setting Gain
If the wave is too small or clipped, you can select [
>>
] and then press [
Gain
] repeatedly in Manual Defibrillation mode to
change its size.
There are altogether 7 options, namely [
Auto
], [
0.125
], [
0.25
], [
0.5
], [
1
], [
2
] and [
4
]. When [
Gain
] is set to [
Auto
],
the system selects the most appropriate gain for the current waveform.
7.5.3
Choosing AHA or IEC Lead Placement
1. Press
the
Power On/Off
button, and then select
→
[
Config.
]
→
[
Config. Edit
]
→
enter the required password.
2.
In the [
Config. Edit
] menu, select [
ECG Setup
→
]
[
ECG Standard
], and then select [
AHA
] or [
IEC
] according to the
standard that is applied to your hospital.
7.5.4
Setting Filter Mode
When monitoring ECG through ECG lead set, filter mode is displayed above the ECG waveform. To change the filter
mode:
1. Press
the
Power On/Off
button, and then select
→
[
Config.
]
→
[
Config. Edit
]
→
enter the required password.
2.
In the [
Config. Edit
] menu, select [
ECG Setup
→
]
[
ECG Bandwidth
], and then select [
Therapy
] or [
Monitor
].
Summary of Contents for BeneHeart D1
Page 1: ...BeneHeart D1 Automated External Defibrillator Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 42: ...6 6 FOR YOUR NOTES...
Page 50: ...7 8 FOR YOUR NOTES...
Page 58: ...9 6 FOR YOUR NOTES...
Page 86: ...B 4 FOR YOUR NOTES...
Page 88: ...C 2 FOR YOUR NOTES...
Page 92: ...D 4 FOR YOUR NOTES...
Page 96: ...E 4 FOR YOUR NOTES...
Page 100: ...F 4 FOR YOUR NOTES...
Page 102: ...G 2 FOR YOUR NOTES...
Page 104: ......
Page 105: ......
Page 106: ...P N 046 004673 00 3 0...