Arrhythmia Analysis
Arrhythmia analysis provides information about your patient’s condition, including heart rate and arrhythmia alarms.
In the process of ECG monitoring, if any arrhythmia event is detected, corresponding arrhythmia alarms are reported
according to the alarm level.
NOTE
Arrhythmia analysis program is intended to detect ventricular arrhythmias. It is not designed to detect
atrial or supraventricular arrhythmias. It may incorrectly identify the presence or absence of an
arrhythmia. Therefore, a physician must analyze the arrhythmia information with other clinical findings.
7.6.1
Understanding the Arrhythmia Events
Arrhythmia event
Description
Category
Asystole
No QRS complex for 4 consecutive seconds (in absence of ventricular fibrillation
or chaotic signals).
Lethal
arrhythmia
Shockable rhythum
A fibrillatory wave for 4 consecutive seconds.
A dominant rhythm of adjacent Vs and a HR > the V-Tach Heart Rate Limit.
Vtac
The consecutive PVCs ≥ Vtac PVCs limit, and the HR ≥ the Vtac rate limit.
Vent. Brady
The consecutive PVCs ≥ the Vbrd threshold and the ventricular HR < the Vbrd
Rate threshold.
Extreme Tachy
The heart rate is equal to or greater than the extreme tachycardia limit.
Extreme Brady
The heart rate is equal to or less than the extreme bradycardia limit.
PVCs/min
PVCs/min exceeds high limit
Nonlethal
arrhythmia
PNP
No pace pulse detected for 1.75 x average R-to-R intervals following a QRS
complex (for paced patients only).
PNC
No QRS complex detected for 300 milliseconds following a pace pulse (for
paced patients only).
PVC
One PVC detected in normal heartbeats.
Couplet
Paired PVCs are detected.
VT>2
More than 2 consecutive PVCs within the last minute.
Bigeminy
A dominant rhythm of N, V,N, V, N, V.
Trigeminy
A dominant rhythm of N, N, V,N, N, V, N, N, V.
R ON T
R on T detected in normal heartbeats.
Brady
The average heart rate is equal to or less than 60 bpm.
Tachy
The average heart rate is equal to or greater than 100 bpm.
Vent. Rhythm
The consecutive PVCs ≥ the Vbrd PVCs limit, and the HR is ≥ Vbrd Rate limit but
< the Vtac Rate limit.
Multif. PVC
Multiform PVCs detected in Multif. PVC's Window (which is adjustable).
Summary of Contents for BeneHeart D1
Page 1: ...BeneHeart D1 Automated External Defibrillator Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 42: ...6 6 FOR YOUR NOTES...
Page 50: ...7 8 FOR YOUR NOTES...
Page 58: ...9 6 FOR YOUR NOTES...
Page 86: ...B 4 FOR YOUR NOTES...
Page 88: ...C 2 FOR YOUR NOTES...
Page 92: ...D 4 FOR YOUR NOTES...
Page 96: ...E 4 FOR YOUR NOTES...
Page 100: ...F 4 FOR YOUR NOTES...
Page 102: ...G 2 FOR YOUR NOTES...
Page 104: ......
Page 105: ......
Page 106: ...P N 046 004673 00 3 0...