3-4
3.2.2.2
Selecting System Language
In the [
General Setup
] menu, select [
Language
] to set the system language, which refers to the language of messages,
menus, and audio prompts and so on.
If the system language is set to a non-English language, you can also set [
Bilingual Option
] in the [
General Setup
]
menu. When [
Bilingual Option
] is set to [
On
], the text prompts in AED mode are displayed in English and the set system
language. And for Public, you can press the Audio Language softkey to switch the language of audio prompts. For
details about the softkey, refer to
2.5 Soft Key Symbols
.
For Pro, [
Bilingual Option
] is disabled when [
ECG Display
] is set to [
On
].
3.2.2.3
Setting Default Startup Mode (For Pro Only)
For Pro, in the [
General Setup
] menu, select [
Default Startup Mode
] and set the default startup mode to:
[
AED
]:
the equipment enters AED mode by default after startup; or,
[
Manual
]: the equipment enters Manual Defib mode by default after startup.
3.2.3
Turning off the Equipment
To turn off the equipment, follow this procedure:
1.
Confirm that the patient monitoring or therapy is completed.
2.
Disconnect the patient cables and sensors from the patient.
3. Press
the
Power ON/OFF
button and the “Select an option” window is displayed.
4. Press
the soft key to shut down the equipment.
3.2.4
Auto Restoring to Last Configuration
During operation, you may make changes to some settings. However, these changes may not be saved as user
configuration. To prevent the changes from losing in case of sudden power failure, the equipment saves the settings in
real time. The saved settings are the latest configuration. In case of power failure, the equipment loads the latest
configuration if restarts within 60 seconds; it loads the user configuration if restarts 120 seconds later after the power
failure; it may load either the latest configuration or the user configuration if restarts between 60 and 120 seconds after
the power failure.
3.3
Post Use Procedure
After the equipment has been used on a patient, the unit shall be cleaned as described in
11 Care and Cleaning
. Then
follow the procedure below to prepare the equipment for next use:
1.
Connect a new pads package to the equipment as described in
3.1.4 Connecting Electrode Pads
.
2.
Perform a user test as described in
12.3.5 User Test
. Check the test result and make sure all test items are passed.
3.
Turn off the equipment.
Summary of Contents for BeneHeart D1
Page 1: ...BeneHeart D1 Automated External Defibrillator Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 42: ...6 6 FOR YOUR NOTES...
Page 50: ...7 8 FOR YOUR NOTES...
Page 58: ...9 6 FOR YOUR NOTES...
Page 86: ...B 4 FOR YOUR NOTES...
Page 88: ...C 2 FOR YOUR NOTES...
Page 92: ...D 4 FOR YOUR NOTES...
Page 96: ...E 4 FOR YOUR NOTES...
Page 100: ...F 4 FOR YOUR NOTES...
Page 102: ...G 2 FOR YOUR NOTES...
Page 104: ......
Page 105: ......
Page 106: ...P N 046 004673 00 3 0...