4-3
4.4
Alarm Tone Configuration
4.4.1
Changing the Alarm Volume
1.
If the equipment is on, press the Power On/Off button and then select
→
[
Config.
→
]
[
Config. Edit
→
]
enter
the required password
→
[
Alarm
].
2. Set
[
Alm Volume
] to any of the following:
[
High
]: the alarm volume is set to the highest level.
[
Med
]: the alarm volume is set to a medium level.
[
Low
]: the alarm volume is set to a lower level.
[
Off
]: the alarm sound is disabled.
NOTE
The alarm volume for special system alarms is always high and not user-adjustable.
4.4.2
Pausing Alarm Sounds
You can press the [
Silence
] softkey to temporarily disable alarm tones. In this case, the symbol
will be displayed in
the sound symbol area indicating all alarm sounds are silenced temporarily. In the audio paused status, all alarm
indicators except audible alarm tones works properly. You can press [
Silence
] again to restore alarm sounds.
If new alarms are triggered, alarm sounds restore to normal automatically.
NOTE
The alarm volume for special system alarms cannot be paused.
4.4.3
Switching Off Alarm Sounds
1.
If the equipment is on, press the Power On/Off button and then select
→
[
Config.
→
]
[
Config. Edit
→
]
enter
the required password
→
[
Alarm
].
2. Set
[
Alm Volume
] to [
Off
] to switch off the alarm sounds.
In the audio off status,
appears in the sound symbol area. In this case, all alarm indicators except audible alarm
tones works properly. To resume the alarm sounds, set [
Alm Volume
] to [
High
], [
Med
] or [
Low
].
When alarms or alarm sounds are turned off, the equipment can give a reminder tone of a single beep every 60 seconds.
The volume for reminder tone is set to a fixed level and not user-adjustable.
Summary of Contents for BeneHeart D1
Page 1: ...BeneHeart D1 Automated External Defibrillator Operator s Manual...
Page 2: ......
Page 8: ...VI FOR YOUR NOTES...
Page 42: ...6 6 FOR YOUR NOTES...
Page 50: ...7 8 FOR YOUR NOTES...
Page 58: ...9 6 FOR YOUR NOTES...
Page 86: ...B 4 FOR YOUR NOTES...
Page 88: ...C 2 FOR YOUR NOTES...
Page 92: ...D 4 FOR YOUR NOTES...
Page 96: ...E 4 FOR YOUR NOTES...
Page 100: ...F 4 FOR YOUR NOTES...
Page 102: ...G 2 FOR YOUR NOTES...
Page 104: ......
Page 105: ......
Page 106: ...P N 046 004673 00 3 0...