5 - 2
5.3
Alarm Solutions
WARNING
•
When an alarm occurs, check the pump’s status and handle the alarm as soon
as possible. If the alarms do not conform with the actual situation, contact
your service personnel.
NOTE
•
The pump stops infusion when a high priority alarm is triggered.
•
The pump continues infusion when a low priority alarm is triggered.
•
The pump stops infusion after the first Battery Depleted alarm occurs, and
the shutdown delay is at least three minutes.
5.4
Occlusion Alarm
NOTE
•
If this pump is running at 0.1ml/h, and respectively configure the occlusion
pressure alarm limit to the lowest level and highest level, the occlusion alarm
delay time may reach up to one hour and 20.5 hours. Adjust the pressure
limit to a lower level, or use the syringe pump for a low rate infusion.