A - 6
A.5
Occlusion Alarm Delay and Bolus Volume
Test conditions:
■
Infusion set brand: B.Braun Intrafix Primeline )
■
Anti-bolus: On
■
Test temperature: 20
°C
±2
°C
■
Infusion line length: 1 meter
WARNING
•
Occlusion alarm pressure, alarm delays and bolus volume may vary
depending on test conditions, temperature and tube length.
A.6
Infusion Accuracy Graphs
A.6.1
Infusion Accuracy at 1 ml/h
Rate (ml/h)
Occlusion alarm delay time (hh: mm: ss)
High occlusion alarm pressure
level
Low occlusion alarm pressure
level
1
<
02:01:00
<
00:04:37
25
<
00:03:25
<
00:00:30
Rate
Bolus volume after occlusion (ml)
High occlusion alarm pressure
level
Low occlusion alarm pressure
level
25ml/h
<
0.18
<
0.18