B - 7
TABLE EMC-5- Test specifications and minimum distances
Recommended separation distances between portable and mobile RF
communications equipment and the device
The system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended
below, according to the maximum output power of the communications equipment.
Portable and mobile radio communications equipment (e.g. two-way radio, cellular/
cordless telephones and similar equipment) should be used no closer to any part of this
system, including cables, than determined according to the following method:
Test
frequency
(MHz)
Band
(MHz)
Service
Modulati
on
Maximu
m power
(W)
Distance
(m)
Immunit
y test
level (V/
m)
385
380 - 390
TETRA
400
Pulse
modulati
on
18Hz
1.8
0.3
27
450
430 -470
GMRS
460
FRS 460
FM ±5
kHz
deviation
1 kHz sine
2
0.3
28
710
704 - 787
LTE Band
13,17
Pulse
modulati
on
217 Hz
0.2
0.3
9
745
780
810
800 - 960
GSM 800/
900, tetra
800, iDEN
820,
CDMA
850, LTE
Band 5
Pulse
modulati
on
18 Hz
2
0.3
28
870
930
Summary of Contents for BeneFusion uVP
Page 1: ...BeneFusion uVP BeneFusion uVP ex Infusion Pump Instruction for Use ...
Page 2: ......
Page 12: ...1 4 This page intentionally left blank ...
Page 18: ...3 4 This page intentionally left blank ...
Page 26: ...6 2 This page intentionally left blank ...
Page 30: ...8 2 This page intentionally left blank ...
Page 32: ...9 2 This page intentionally left blank ...
Page 36: ...10 4 This page intentionally left blank ...
Page 38: ...11 2 This page intentionally left blank ...
Page 40: ...12 2 This page intentionally left blank ...
Page 60: ...B 10 This page intentionally left blank ...
Page 61: ......
Page 62: ...KF H 046 024481 00 4 0 ...