4 - 3
•
If the infusion set type is Nutrition or the selected drug type is Enteral
Nutrition, “Air detection will be closed” is always displayed in the system
information area, “[Enteral feeding] Air detection will be closed” is displayed
before starting the infusion or bolus infusion.
•
The infusion could not be started when the door is open.
•
Monitor the connection of infusion set, tubing, pump and patient, and the
drug information on a regular basis. Start infusion according to the
instructions in this manual.
4.5
Bolus Infusion
NOTE
•
The delivered bolus volume will be added to the total infusion volume and
subtracted from the volume to be infused (VTBI).
•
The pump gives a beep every time a 0.5 ml bolus volume is infused.
4.6
Setting Keep Vein Open (KVO) Rate
NOTE
•
If the KVO rate is greater than the infusion rate, the pump will continue to
infuse at the set infusion rate.
•
The pump runs for 30 minutes at a KVO rate. At the completion of the KVO
infusion, the pump stops infusion, and gives a KVO Finish alarm.
•
The volume used during KVO infusion will be added to the total infusion
volume.
4.7
Replacing the IV Container
NOTE
•
Check the infusion status after replacing the IV container. If the pump is
running a KVO infusion, reconfigure the infusion parameters before
restarting the infusion.
Summary of Contents for BeneFusion uVP
Page 1: ...BeneFusion uVP BeneFusion uVP ex Infusion Pump Instruction for Use ...
Page 2: ......
Page 12: ...1 4 This page intentionally left blank ...
Page 18: ...3 4 This page intentionally left blank ...
Page 26: ...6 2 This page intentionally left blank ...
Page 30: ...8 2 This page intentionally left blank ...
Page 32: ...9 2 This page intentionally left blank ...
Page 36: ...10 4 This page intentionally left blank ...
Page 38: ...11 2 This page intentionally left blank ...
Page 40: ...12 2 This page intentionally left blank ...
Page 60: ...B 10 This page intentionally left blank ...
Page 61: ......
Page 62: ...KF H 046 024481 00 4 0 ...