A - 4
Set range of the infusion rate/
purge rate/bolus rate
0.1ml/h to 2000ml/h
Minimum resolution:
0.01ml/h (0.1 to 99.99ml/h)
0.1ml/h (100.0 to 999.9ml/h)
1ml/h (1000 to 2000ml/h)
Range of the infusion rate for micro-infusion mode:
0.1ml/h to 100ml/h
Set rang of drop rate
(1 to 400) d/min
Occlusion pressure 50mmHg
to
1125mmHg
Resolution: 1mmHg
The maximum occlusion pressure is 1350mmHg.
Occlusion pressure tolerance
50mmHg to 149mmHg:
≤
±75mmHg (operating at a
rate
≤
100 ml/h)
150mmHg to 1125mmHg:
≤
±20% or
≤
±125mmHg,
whichever is greater
Maximum volume (under
single fault conditions)
≤
0.5ml
KVO rate
0.1 to 5.0ml/h
Minimum resolution: 0.01ml/h
Time set range
00:00:01 to 99:59:59 (h:min:sec)
VTBI set range
0.1 to 9999.99 ml
Minimum resolution: 0.01ml
Micro-infusion mode: 0.1 to 1000ml
Weight set range
0.1 to 499.0 kg/0.2 to1100.1 lb
Drug Amt. set range
0.001 to 99999
Drug Amt. unit set range
ng, μg, mg, g, mU, U, kU, EU, mmol, mol, mcal, cal, kcal,
mEq
Volume set range in Dose Time
Mode/Dose Mode
0.1 to 9999.99ml
Conc. set range
0.001 to 9999.99
Conc. unit set range
ng/ml, μg/ml, mg/ml, g/ml, mU/ml, U/ml, kU/ml, EU/ml,
mmol/ml, mol/ml, mcal/ml, cal/ml, kcal/ml, mEq/ml
Dose Rate set range
0.001 to 99999
Patient management
Discharging/admitting a patient,
Editing/exporting/importing patient information
Summary of Contents for BeneFusion uVP
Page 1: ...BeneFusion uVP BeneFusion uVP ex Infusion Pump Instruction for Use ...
Page 2: ......
Page 12: ...1 4 This page intentionally left blank ...
Page 18: ...3 4 This page intentionally left blank ...
Page 26: ...6 2 This page intentionally left blank ...
Page 30: ...8 2 This page intentionally left blank ...
Page 32: ...9 2 This page intentionally left blank ...
Page 36: ...10 4 This page intentionally left blank ...
Page 38: ...11 2 This page intentionally left blank ...
Page 40: ...12 2 This page intentionally left blank ...
Page 60: ...B 10 This page intentionally left blank ...
Page 61: ......
Page 62: ...KF H 046 024481 00 4 0 ...