M
AINTENANCE
- 40 -
F
LUSH
M
ODE
The Tank Outlet Valve should be left open
during FLUSH mode.
The system will now go into FLUSH mode for
15 minutes (H). This is done to empty the
SmartPak of air and hydrate the material inside.
When FLUSH mode is finished, the system will
go into FILLING TANK mode automatically
(I).
R
INSING THE
S
MART
P
AK
Let FILLING TANK mode run for 15 minutes
with the Tank Outlet Valve open. This will rinse
the purification media inside the SmartPak.
Close the Tank Outlet Valve (J). The tank will
start to fill up with water. It could take
approximately 1. 5 to 2 hours to fill the tank.
When it is full, the Display will show the Tank
Level filled to the 100% level.
NOTE: for sensitive applications, it is
recommended to leave the system in FILLING
TANK overnight to ensure complete rinsing of the
RO membrane.
H
I
J
ATTENTION