I
NSTALLATION
- 19 -
4-9
I
NSTALLATION AND
R
INSING OF THE
F
INAL
F
ILTER
Obtain a Final Filter. It can be the Millipak Express 20 or the BioPak Ultrafiltration Cartridge.
Remove the Tygon Tubing and the Barbed Fitting from the POU Dispenser.
Screw the Final Filter onto the end of the POU Dispenser.
The Final Filter should be turned until it is
hand tight (A). Do not over tighten the Final Filter.
Do not use white tape on the threads of the Final
Filter. The POU Dispenser has an O-ring inside
which provides a watertight seal.
Press the Dispensing Button once. This will bring the system into DISPENSING mode.
(If your Final Filter is a Millipak)
Locate the air vent valve on the top side of the Millipak. Open this
slowly but do not remove it from the Millipak (B). Allow any air on the clear side of the Millipak to be
vented out.
Air will not pass through the membrane filter in
the Millipak. If there is trapped air in the
Millipak, then a reduction in Product Water
flowrate can result.
Dispense about 1 Litre of water. Make sure all air is purged out.
Press the Dispensing Button once again. The system will finish dispensing water.
Leave the system in FILLING TANK mode.
It is highly recommended not to put Tygon
Tubing or any other type of tubing on the end of
the Final Filter. This can compromise the Product
Water quality (bacteria can grow in the Tygon
Tubing).
The Water System is now ready for use.
A
B
ATTENTION
ATTENTION
ATTENTION