I
NSTALLATION
- 18 -
4-7
P
URGING
A
IR FROM THE
S
YSTEM
At this time you should have installed the
SmartPak, Barbed Fitting and the Tygon
Tubing. Air trapped in the SmartPak should now
be purged from the system.
Verify that you have a full tank of RO water by
viewing the Tank Level display (A).
Press the Dispensing Button once to put the
system into DISPENSING mode (B).
Dispense the full tank of water from the system.
The system will go into FILLING TANK mode
when the tank level is below the 60% level (C).
When the tank level reaches the minimum level,
the system will finish dispensing water.
4-8
H
YDRATING THE
S
YSTEM
At this time, the Final Filter is not installed.
Leave the system overnight or for several hours
in FILLING TANK mode or PRE OPERATE
mode (see
Section 5-3 Operating Modes
). The
system will regularly recirculate water and rinse
off purification media inside the SmartPak. Do
not leave the system in STANDBY mode.
(The next morning) Press the Dispensing Button
once to put the system into DISPENSING mode.
Allow about 1 Litre of water to be dispensed
from the system.
Press the Dispensing Button once again. The
system will finish dispensing water.
Remove the Tygon Tubing and the Barbed
Fitting from the POU Dispenser.
The Final Filter needs to be installed. See
Section 4-9
Installation and Rinsing of the Final Filter.
A
B
C