I
NSTALLATION
- 20 -
4-10
H
OW TO
C
ALIBRATE THE
F
LOWRATE FROM THE
D
IRECT
-Q
(F02)
Before calibrating the Product flowrate from the Direct-Q, you will need a 1 Litre graduated
cylinder to measure the total volume of water that will be dispensed. The Final Filter should be
installed.
1.
Press the Main and
“-“ Buttons together
to enter the menu.
The Display will
show F01.
2.
Press the Main
Button once. The
Display will show
F02 and the
60 second timer. The
graduated cylinder
will be blinking.
3.
Press the Dispensing
Button once. The
system will dispense
water for 60 seconds.
The Display will
show the timer
counting down.
NOTE: If you press the Dispensing Button again or if the Tank Level display is at the 10%
level before 60 seconds have elapsed, then the system will stop dispensing water. Calibration
of the flowrate is cancelled.
CONTINUED ON NEXT PAGE