MacroPATH and MacroPATH Basic QX Series
MM161-001
– Operator Manual
54
High definition
this function is not necessary for the camera (it is developed for previous camera models).
Enabled shared folder management
Shared folder management is useful when several MacroPATH stations are working on the same files. In
fact, with this function enabled, when a case is opened by a station and another MacroPATH station tries to
open it, an error message is displayed, informing that the case is already opened by another System and
displaying its IP address and when the case has been opened. An example is shown below.
Before
using
this
function,
please
contact
your
Technical
Department
and
3.6.5. Camera Settings
Click the “Camera” button to enter the camera settings page
Select the required
camera (“Cam1” or “Cam2”, “Cam3”, “Cam4” and “Cam5”
depending on how many there are).
Click on
Camera short name
to edit the camera name.
Select:
•
Trust external calibration
: MacroPATH to use the calibration factors of previous files and not those
of the software.
•
Force calibration
: MacroPATH will require calibration to be carried out on every new image (helpful
when changing capture camera height).
•
Auto measure:
function not available.
•
Vertical Flip Image:
the image is flipped along the vertical axis.
•
Horizontal Flip
: the image is flipped along the horizontal axis.
•
Priority
: If two or more cameras are connected to MacroPATH, you must choose the selection
sequence.
It is essential to set
the priority when working with multiple cameras. ‘Live’ shows the
buttons for the various connected cameras, following the set order of priority.