MacroPATH and MacroPATH Basic QX Series
MM161-001
– Operator Manual
44
3.5.1. Tools
Sizing
Click and hold down the
Sizing
icon with the mouse or touch pen.
The tool bar circled in red will appear.
Click the “Measure length” icon to draw lines; the system calculates the length of the line.
Click the “Sizing - Area” icon to draw an area on the sample; the system calculates the drawn area.
Click the “Sizing - Free” icon to draw freehand lines; the system calculates the length of the line.
The Software needs to be calibrated when the unit is installed and every time the camera height
changes. See chapter 3.6.6 for the calibration procedures.
MacroPATH
's
measurement
function is strictly linked to the calibration performed by the operator and is not controlled
by the unit's software. That is why the measurements of the sample obtained with
MacroPATH
CANNOT be taken into account for diagnostic purposes. These values are
purely indicative.
Indeed, a ruler must be placed near the sample (see image below) or a gauge must be used to obtain
a diagnostic value for the growth of the tumour, the extent of the lesion, etc.
If there is no calibration, the <?> symbol appears on the image.