43
Instruction Manual
8.2.6 Disinfecting the upholstery
Following disinfectants have only slight adverse effects on the uphol-
stery provided that the indicated dosing is observed and that it is ap-
plied appropriately. The listed products are a recommended selec tion.
A complete list of suitable disinfectants (320.8724.02 Upholstery Main-
tenance) can be requested at Mikrona.com / Downloads / Instruction.
Denotation
max.
Manufacturer
Concen-
tration
PlastiSept*
pure
Alpro, DE-St. Georgen
PlastiSept-wipes*
pure
Alpro, DE-St. Georgen
Acrylan
pure
Anticeptica, DE-Pulheim
Bacillol 25
pure
Bode Chemie, DE-Hamburg
Meliseptol Foam fresh
pure
Braun Medical, DE-Melsungen
SprayIn
pure
Deppe, DE-Kempen
SprayOff
pure
Deppe, DE-Kempen
FD 300 top wipes*
pure
Dürr Dental, DE-Bietigheim
FD 312*
1,0%
Dürr Dental, DE-Bietigheim
FD 312 wet wipes*
pure
Dürr Dental, DE-Bietigheim
FD 366 sensitive
pure
Dürr Dental, DE-Bietigheim
FD 366 sensitive wipes
pure
Dürr Dental, DE-Bietigheim
Kanitop-AF*
pure
Kaniedenta, DE-Herford
Köhler Sprühdesinfektion
pure
Köhler, DE-Neckarsulm
Medichem Desinfektionschaum pure
Medichem, DE-Rendsburg
Biosanitizer-S wipes*
pure
Saniswiss, CH-Genève
Cleanisept wipes
pure
Schumacher, DE-Melsungen
Mikrozid sensitive Liquid* pure
Schülke&Mayr, DE-Norderstedt
Mikrozid sensitive- wipes* pure
Schülke&Mayr, DE-Norderstedt
Unisepta Foam- wipes*
pure
Unident, CH-Genève
*alcohol-free disinfectant
Dry the upholstery after disinfecting it, in order to prevent the liquid
from penetrating the upholstery.
8.3
Annual maintenance by the service technician
8.3.1 Safety checks
Medical equipment is designed in such a way that the firs occurence
of a fault does not create a hazard to the safety of patients, users or
other persons.
For that reason it is essential to perform safety tests every 2 years
which aim particulary at detecting electrical faults. This is done by an
authorized service technician from your dental depot, most pracically
together with the work to be performed according to maintenance.
Safety test must also be performed and documented during initial
startup, after extensions and after repair work which might affect the
electrical safety of the system.
The dental treatment unit must not be operated if it has failed to pass
the safety tests!
As a user, please observe country-specific requirements.
The device must be regularly inspected by the operator or a third par-
ty. It is recommended that the annual maintenance work is carried out
by an entity authorized for this purpose by Mikrona or Mikrona’s own
customer service. Worn or defective parts relevant to security must be
repaired with original parts or replaced.
The work steps to be performed as well as the parts which must be
replaced are specified in the document Care 320.8791.11
Additional Information at Mikrona.com / Service