26
ORTHORA 200
Instruction Manual
7.11 Ultrasound Suprasson / Newtron Satelec
General information
For information on technical features as well as on operation and
application of this ultrasound system please consider the specific in-
struction manual which comes enclosed in the instrument packaging.
Please service and handle the hand piece and inserts as specified in
the maintenance instructions of the specific manufacturer.
Description
This ultrasound system is equipped with integrated spray and optimal
light for dental application.
Commissioning
Never attach any inserts or a working tips while the ultrasound system
is running.
Heart pacemakers may be disrupted by high-frequency ultrasonic
oscillations. Therefore we recommend not to use this product with pa-
tients requiring a pacemaker.
Only use instruments in perfect operating condition. Immediately stop
any work if intermittences, extensive vibrations, abnormal heating or
other signs of instrument failure occur.
For use, remove the instrument from the rack. The display shows the
specific instrument as well as the selected intensity level. Start the
instrument by using the function pedal. Change the intensity by using
the function pedal (7.2.3. or 7.2.4).
Sterilizing the ultrasound device
Clean and sterilize the ultrasound device before using it on the next
patient. Remove the working tip before sterilizing the device and sepa-
rate the hand piece from the hose. After each sterilization cycle, imme-
diately remove the device from the sterilizer to prevent corrosion.
Sterilization occurs in an autoclave at 134°C.
Disinfection
For disinfection use a clean damp cloth and disinfect the case surface.
Only use disinfectants recommended by the manufacturer.
Never submerge the device in a disinfectant solution.
Do not insert the device into an ultrasonic bath.
Maintenance
No specific maintenance procedures are intended for this devise.
Under any circumstances avoid using any lubricants as this my cause
irreparable damage to the ultrasound.
Default settings
Water
ca. 40 ml/min
4-function pedal control 09 (ST09) units with serial numbers > OT1.02000
X
_
X
_
Level 1
0 to 2,0 V
0 to 35%
Level 2
2,0 to 3,5 V
35 to 75%
Level 3
3,5 to 5,0 V
75 to 99%
Level 4
0 to 5,0 V
0 to 99%
5-function pedal control 09 (ST09) units with serial numbers > OT1.02000
X
_
Level 1
0 V
0%
Level 2
2,0 V
35%
Level 3
3,5 V
75%
Level 4
5,0 V
99%
5-function pedal control 01 (ST01) units with serial numbers < OT1.01999
Level 1
0 V
Level 2
2,0 V
Level 3
3,5 V
Level 4
5,0 V