
29
Instruction Manual
7.14 3-function syringe M1600
General information
For information on technical features as well as on operation and ap-
plication of this syringe please consider the specific instruction manual
which comes with the instrument. Please handle the hand piece as
specified in the maintenance instructions of the specific manufacturer.
Description
This syringe has been design exclusively for dental application. It ser-
ves to inject air and water, separated or as spray at room temperature.
Commissioning and use
Only use instruments in perfect operating condition. Immediately stop
any work if intermittences or other signs of instrument failure occur.
Injecting cold water: press left button on the handhold.
Injecting cold air: press right button on the handhold.
Injecting cold water-air-mixture (spray): simultaneously press both but-
tons at the syringe head.
Cleaning and sterilizing
Clean and sterilize the case and the cannula before using it on the
next patient. Remove the cannula by pulling and/or remove the entire
case by loosening the case and unscrewing the plug nipples.
Sterilization occurs in an autoclave at 134°C.
Disinfection
For disinfection use a clean damp cloth and disinfect the case surface.
Only use disinfectants recommended by the manufacturer.
Never submerge the device in a disinfectant solution.
Do not insert the device into an ultrasonic bath.
Maintenance
No specific maintenance procedures are intended for this devise.
Under any circumstances avoid using any lubricants as this my cause
irreparable damage to syringe.
Default settings
Air
3.5 bar (12-15 l/min)
Water
ca. 100 ml/min