7.2.4 Instrument operation 5-F
Instrument operation 5-F
Micromotor, turbine and ultrasonic devices are controlled by the
5-function pedal and shown on the display A.
10
Speed level down or intensity level down
11
Speed level up or intensity level up
12
Device on
13
Spray on
14
Chip blower on/off
Water supply spray suppressed
LED on the display shines
14
Micromotor changes direction of rotation when pressing button for
more than 1.5 seconds.
The basic settings can be programmed under the instruments data IN (7.2.6).
Stand-alone 2-function pedal
Micromotor, turbine and ultrasonic devices are optionally controlled by
the stand-alone 2-function pedal and shown on the display A.
The stand-alone 2-function pedal can only be used in parallel connec-
tion with the built-in 5-function pedal.
Instrument operation:
1
Device on
2
Device on, spray on
Speed / intensity of instruments is adjusted by the built-in 5-function
pedal.
Stand-alone 5-function pedal
Micromotor, turbine and ultrasonic devices are controlled by the stand-
alone 5-function pedal and shown on the display A.
The stand-alone 5-function pedal can be used autonomously or in par-
allel connection with the built-in 5-function pedal.
Instrument operation:
1
Speed level down or intensity down
2
Speed level up or intensity up
3
Device on
4
Device on, spray on
5
Chip blower on/off
Water supply spray suppressed
LED on the display shines
5
Micromotor change direction of rotation when pressing button for
more than 1.5 seconds.
The basic settings can be programmed under the instruments data IN (7.2.6).
1 2
2
5
1
3
4
With the operation over the 5-function pedal the speed/intensity is
emitted constantly within the 4 selectable levels. The setting of the
speed/intensity within the levels is adjusted in the electronics (ST01)
or in the programming mode (ST09)
Display functions
2
5
1
3
4
Active instruments and their functions are shown on the display.
B Display of special functions
Instrument display
Active instrument is displayed.
C Ultrasonic device
D Turbine (T)
E Micromotor (M)
F Speed level
Speed and intensity can be selected from 4 levels.
Each level per instrument can be adjusted by a service technician
to the individual needs of the user.
Direction of rotation of micromotor:
Clockwise rotation LED shines
Counter-clockwise rotation LED blinks
G Chip blower (C) or , blower air (LED shines)