8. Maintenance

8.1 

Daily maintenance

Operational safety and utilizability of the device are directly connected 
to the recommended care and maintenance measures. Regular main-
tenance and care are therefore indispensable. If there are any indica-
tions that a risk for the patient or operator could arise from abrasion 
of components or from a technical error, the device is to be inspected 
immediately by an authorized customer service or Mikrona’s customer 
service and/or the fault is to be cleared.

Check the function of the magnetic main water valve every day. If 
the magnetic valve for main water supply is working, only little water 
should run from the multifunction syringe when pressing the specific 
button after switching off the main switch (7.1 Main switch).

Take care that no water enters the device when cleaning as this may 
cause errors.

Clean and disinfect the surface of the device regularly, e.g. with Dürr 
Surface Disinfection FD 322 / FD 350.

8.1.2  Daily maintenance of the spittoon

The spittoon is tightly connected to the unit by adapter.

Clean and disinfect the surface of the spittoon regularly, e.g. with Dürr 
Surface Disinfection FD 322 / FD 350.

It is recommended to regularly rinse the spittoon with tap water 
throughout the day, for example to aspirate a glass of water through 
the drain and the spittoon after each patient

It is recommended to disinfect and clean the spittoon at least once a 
day with Orotol ultra. Application once a day: In the evening after the 
last patient. Application twice a day: Before the lunch break and in the 
evening.

After carrying out the disinfection, it is very important to observe the 
residence time (DGHM: 1 h) before continuing to work on the corre-
sponding unit. The same applies to flushing.

In order to create an efficient solution with Orotol ultra (10g/liter) dis-
solve it entirely in lukewarm water right before use. It is not possible 
to keep the solution in stock as it is only durable for a short period of 
time.

Never clean the spittoon incl. the rinsing basin with a household clea-
ner. Due to potentially appearing foaming, neutralizing tensides and 
insoluble abrasives the aspiration system may become disabled. For 
in-between-cleanings of the drain, spittoon and rinsing basin use Oro-
tol MD 550 or Orotol MD 555.

Avoid under any circumstances to uncontrollably mix any products – 
incl. disinfectants – by means of the spittoon (This may result in dama-
ge of the aspiration system).

8.1.1  Hygiene recommendations for  

water-bearing systems

Before starting a new day of treatment, rinse all extraction points, such 
as glass filler, turbine, micromotor, tartar remover and multi-function 
syringe (without instruments attached) for two minutes with water. 

This 

significantly helps to reduce micro-bacterial accumulations caused by 
stagnation.

It is further recommended to rinse used extraction points for 20 se-
conds with water after each patient and at the end of the day.

8.1.3  Cleaning and Disinfecting the spittoon valve unit

Press button for spittoon rinse.
Pour an amount of disinfectant into the spittoon, e.g. Orotol Plus.
Keep pressing button for spittoon valve until the disinfecting agent has 
been completely sucked up.

Do not use foaming or aggressive cleaning agents.

Summary of Contents for ORTHORA 200

Page 1: ...320 8710 02 2103 ORTHORA 200 INSTRUCTION MANUAL 2103...

Page 2: ...tallation 13 6 Connections 13 6 1 Electrical connection 13 6 2 Aspiration connection 13 6 3 Air connection 13 6 4 Water connection 13 6 5 Transport and storage conditions 13 6 6 Operating conditions 1...

Page 3: ...spittoon 39 8 1 3 Cleaning and Disinfecting the spittoon valve unit 39 8 1 4 Daily maintenance of the rinsing basin 40 8 1 5 Daily maintenance of the cup holder 40 8 1 6 Daily maintenance of the bubb...

Page 4: ...ate within 10 days after the delivery of the treat ment unit 1 Notes 1 1 Warranty Liability 1 2 Conformity assessment The product ORTHORA 200 has undergone a conformity assessment in accordance with t...

Page 5: ...ould take the following safety measures to eliminate remaining danger Mikrona s products are constructed in compliance with the corre sponding safety related regulations and meet the legal stipulation...

Page 6: ...ols Important information in the operating instructions concerning the safety of individuals and the device are highlighted with the following terms and symbols The symbol caution labels information a...

Page 7: ...stomer service Technical support for the treatment unit is provided by qualified ser vice entities specialized dealers Service technicians instructed by Mikrona are continuously trained and are famili...

Page 8: ...The loading capacity of the suspended table depends on the prestressed spring package and the instrument tray The maximal loading capacity of the suspended table is 2 7 kg and must not be exceeded Due...

Page 9: ...tion First class materials distinguish the ORTHORA 200 treatment unit and guarantee low maintenance and durability Surfaces are smooth and easy to clean and disinfect Optionally a DVGW certified auton...

Page 10: ...nd as DrawerCart with and without instrument adapters Optionally the ORTHOCART is also available with a swiveling table for two standard trays and an instru ment rack for up to 6 instruments The instr...

Page 11: ...4 stage adjustable intensity Abrasive hand piece Yes Polymerization lamp Yes Storage possibilities Yes equipment tray ORTHODESK swiveling table Air syringe Yes 11 Multifunction syringe Cold water Yes...

Page 12: ...4 2 Dimensions max working height 750 mm min working height 580 mm Treatment positions 2 freely programmable positions Weight 110 150 kg depending on equipment Footprint 2600 cm2 Floor loading 1 88 N...

Page 13: ...pe 1 10 mm per meter Water filtration on device 80 m Ph value 7 2 7 8 Water hardness 7 11 dH 13 19 fH If the water is harder than indicated it is necessary to install a water softening device If the w...

Page 14: ...in parallel with the built in instrument pedal stand alone 2 function pedal or stand alone 5 funtion pedal 7 2 4 stand alone 1 function pedal or stand alone 4 function pedal DYN 7 2 3 7 2 1 Chair cont...

Page 15: ...per level in mode IS Speed intensity of instruments is adjusted by the built in 4 function pedal DYN Stand alone 4 function pedal DYN dynamic Micromotor turbine and ultrasonic devices are optionally...

Page 16: ...the display A The stand alone 5 function pedal can be used autonomously or in par allel connection with the built in 5 function pedal Instrument operation 1 Speed level down or intensity down 2 Speed...

Page 17: ...he maximum The default setting can be individually adjusted and saved by the user Activate all instruments controlled by the 5 function pedal one after another and set them at the desired parameters T...

Page 18: ...nd CL are service func tions They may only be changed by trained staff Individual basic adjustment of the instruments data The chosen speed is saved TA 1 TA 2 TA 3 TA 4 SETUP ENTER Setup Setup Cu DN C...

Page 19: ...pedal for instrument operation optional 5 Main switch 7 5 Spittoon I Bubbler on K Glass filler on off The filling process can be interrupted at all time by slightly pres sing the on off button Fillin...

Page 20: ...removing a aspi ration hose the aspiration system will be started by an electrical si gnal The hose holder with its aspiration hoses and hand pieces works with dry and wet aspiration systems Any othe...

Page 21: ...or 1 2 Micromotor 2 3 Ultrasonic 4 Turbine 1 5 Turbine 2 6 1 H Micromotor 1 2 H Micromotor 2 3 H Ultrasonic 4 H Turbine 1 5 H Turbine 2 6 H 1 C Micromotor 1 2 C Micromotor 2 3 C Ultrasonic 4 C Turbine...

Page 22: ...direction and switch on the spray by using the function pedal 7 2 3 or 7 2 4 Sterilizing the micromotor hand piece The micromotor hand piece which is not included in the scope of de livery of Mikrona...

Page 23: ...ay by using the function pedal 7 2 3 or 7 2 4 Sterilizing the micromotor hand piece The micromotor hand piece which is not included in the scope of de livery of Mikrona Technologie AG should be servic...

Page 24: ...g the function pedal 7 2 3 or 7 2 4 Sterilizing the turbine hand piece The turbine hand piece which is not included in the scope of delivery of Mikrona Technologie AG should be serviced and handled ac...

Page 25: ...intensity by using the function pedal 7 2 3 or 7 2 4 Sterilizing the ultrasound device Clean and sterilize the ultrasound device before using it on the next patient Remove the working tip before ster...

Page 26: ...pedal Change the intensity by using the function pedal 7 2 3 or 7 2 4 Sterilizing the ultrasound device Clean and sterilize the ultrasound device before using it on the next patient Remove the workin...

Page 27: ...njecting air press right button on the handhold Injecting water air mixture spray simultaneously press both buttons on the handhold Cleaning and sterilizing Clean and sterilize the case jacket and the...

Page 28: ...ss the left button on the handhold Injecting warm air Turn the switch on the bottom of the handhold to the right When the green LED diode shines press the right button on the handhold Injecting warm w...

Page 29: ...n the handhold Injecting cold air press right button on the handhold Injecting cold water air mixture spray simultaneously press both but tons at the syringe head Cleaning and sterilizing Clean and st...

Page 30: ...tton on the handhold Injecting water press right button on the handhold Injecting water air mixture spray simultaneously press both buttons on the handhold Cleaning and sterilizing Clean and sterilize...

Page 31: ...tion of dental material Commissioning and use Only use instruments in perfect operating condition Immediately stop any work if intermittences abnormal heating or other signs of instru ment failure occ...

Page 32: ...g and maintenance instruc tions which come with the lamp Operation the halogen lamp low light intensity high light intensity Operation LED ON OFF turn on the lamp L control light intensity LUX K color...

Page 33: ...p is activated and let the halogen lamp and the reflector cool down sufficiently before remo ving the protective cover Unscrew the central attachment bolts of the protective cover counter clockwise an...

Page 34: ...ver while the lamp is activated and let the halogen lamp and the reflector cool down sufficiently before remo ving the protective cover Unscrew the central attachment bolts of the protective cover cou...

Page 35: ...tive cover is made of acrylic glass and easy to clean Clean stained surfaces by using warm water with a non abrasive household cleaner and a soft tissue or cloth cotton Avoid under any circum stances...

Page 36: ...steel wool or Scotch Scotch Brite Hand Pad Nr 7447 rubbing in direction of the surface s grain Disinfection Use a moist cloth and disinfect the surface with a standard disinfectant e g D rr Surface Di...

Page 37: ...th a disinfectant solu tion e g Bilpron A Depressurize the water bottle toggle switch on 2 OFF B Unscrew the water bottle clockwise C Empty the water bottle from tap drinking water D Fill the water bo...

Page 38: ...sui table for a medical assistant Re Forma Lite The Re Forma Lite is the simplest version among the three available It offers a wide circular seat adjustable in height Re Forma Hi The Re Forma Hi work...

Page 39: ...lication twice a day Before the lunch break and in the evening After carrying out the disinfection it is very important to observe the residence time DGHM 1 h before continuing to work on the corre sp...

Page 40: ...nor nitro thinners or synthetic resin thinners Use a standard surface disinfectant to disinfect the cup holder Regu larly clean and disinfect the surface of the cup holder e g with D rr Surface Disin...

Page 41: ...nge and replace it The filter has been desi gned for disposable use and is not suitable for autoclavation It is heat resistant up to approx 60 C Never work without filter otherwise solid particles may...

Page 42: ...cted for breaks and changed if necessary Use a standard surface disinfectant to disinfect the hoses Regularly clean and disinfect the surface of the hose holder e g with D rr Sur face Disinfection FD...

Page 43: ...E Norderstedt Unisepta Foam wipes pure Unident CH Gen ve alcohol free disinfectant Dry the upholstery after disinfecting it in order to prevent the liquid from penetrating the upholstery 8 3 Annual ma...

Page 44: ...quality and replace water filter if necessary Replace water valve and tappet 320 3069 01 Replace water valve and tappet 320 3069 01 Replace water valve and tappet 320 3069 01 Open bleed screw Adjust p...

Page 45: ...er valve and tappet 320 3069 01 Open bleed screw Adjust pressure values A1 A2 Replace tappet to water valve 320 3016 01 Check NC or NO valve Replace control valve 320 7010 01 Replace magnetic valve 32...

Page 46: ...in question Retighten screw coupling hand tight Check water quality and replace water filter if necessary Replace water valve and tappet 320 3069 01 Replace water valve and tappet 320 3069 01 Replace...

Page 47: ...ve micro switch soldered joint on pe dal 4 F2 Defective micro switch soldered joint on pe dal 4 F3 Defective micro switch soldered joint on pe dal 3 F2 Defective top end switch for backrest drive Defe...

Page 48: ...or valve Loose connection between distributor hose and tube Soiled filter on E valve Soiled borehole in overflow line Improperly attached cannula Soiling within bellows system Clogged valve within sel...

Page 49: ...seal pipe adapter and cup filling pipe and or replace O rings 902 0080 01 Possible causes Trouble shooting 9 9 Comfort hose holder Declining suction No suction Hand piece slide is out of order or has...

Page 50: ...ittoon valve Remove alginate Insert modified version of drain cover 320 0533 10 Squeaking backrest drive Manual raising of backrest disabled Manual reclining of backrest disabled Squeaking bed drive M...

Page 51: ...may only use components which are listed in the instruction manual or com ponents which have been stated for a specific purpose by Mikrona Technologie AG Safety reliability and function are only provi...

Page 52: ...NA Polaris RAL 9002 RAL 9010 RAL 9016 customer specific finish sensor switch ORTHOLUX 200 ceiling model Rack with aspirator Yes OP suction Saliva suction No Instrument rack max 4 instruments Micromoto...

Page 53: ...ringe LUZZANI Polymerization lamp MINI LED ORTHO Instrument operation Yes No Instrument operation with Electric pedal integrated parallel to OT200 4 Function pedal DYN with display autonomous BOH BOH...

Page 54: ...werCart DrawerCart with instruments adapter Swivel table Yes No Instrument rack max 6 instruments Micromotor MX2 LED Micromotor MC3 IR Micromotor MC3 LED Turbine device 4VLM LUX Ultrasonic PIEZON EMS...

Page 55: ...Micromotor MX MX2 Micromotor MC3 turbine device 4VLM LUX Ultrasonic PIEZON EMS Ultrasonic NEWTRON Satelec 3 Function syringe LUZZANI 6 Function syringe LUZZANI Media supply ORTHODESK Yes only possible...

Page 56: ...MIKRONA TECHNOLOGIE AG Wiesenstrasse 36 8952 Schlieren Switzerland P 41 56 418 45 45 swiss mikrona com www mikrona com...

Reviews: