External dispenser pumps
78
18.2 Electrical connection
IMPORTANT
maximum current drain at the connector strip: 0.5 A.
Connect the external dispenser pump to the 5-pin plug.
Terminals (arranged from top to bottom)
Terminal
Voltage is on
Condition
Restriction
L1
Continuous voltage
N
Rinse (O18)
parallel to solenoid valve (O7)
while the machine is filling.
during
rinsing.
P709 = 0
a)
c)
Wash (O20)
while the circulating pump is running
a)
c), d), e)
Fill / Water
(O17)
while the machine is filling.
P704 = 0
a), b)
parallel to solenoid valve (O7)
while the machine is filling.
during
rinsing.
P704 = 1
a)
c)
a) Hood / door closed
b) Minimum water level in the tank
c) Not in the self-cleaning programme
d) Not while the machine is filling
e) Not in descaling or basic cleaning programme
The parameters P704 and P709 control when a voltage is present at the Fill/Water and Rinse terminals.
Possible settings
P704 Use
0
for dispenser pumps installed outside the machine
1
for dispenser pumps installed outside the machine
2
for dispenser pumps installed outside the machine
3
for internal dispenser pumps
P709 Use
0
for dispenser pumps installed outside the machine
1 not
used
2
for dispenser pumps installed outside the machine
3
for internal dispenser pumps
Summary of Contents for PG 8172
Page 1: ...PG 8172 PG 8172 ECO PG 8172 l PG 8172 ECO ...
Page 3: ......
Page 39: ...Elektroanschluss 36 17 3 Anschlussschema ...
Page 79: ...Electrical connection 76 17 3 Connection diagram ...
Page 119: ...Raccordement électrique 116 17 3 Schéma de raccordement ...
Page 159: ...Collegamento elettrico 156 17 3 Schema di collegamento ...
Page 199: ...Elektrische aansluiting 196 17 3 Aansluitschema ...