62
<Command>
:MEASure:BIN:LIMit <nominal value NR3 >
:MEASure:BIN:LIMit?
Function
:
It sets or queries the BIN LIMIT.
Description
:
Set syntax:MEASure:BIN:LIMit 0.001,100M,1k,1000k
Query syntax:MEASure:BIN:LIMit?
Return data +1.000000E-03, +1.000000E-01, +1.03, +1.06 (Format is in
<NR3>)
<Command>
:MEASure:FILE:LOAD <filename>
Function
:
It opens the meter mode's file.
Description
:
Set syntax:MEASure:FILE:LOAD MICROTEST.( load the file
“MICROTEST”)
<Command>
:MEASure:FILE:LOAD?
Function
:
It queries the meter mode's filename which is being using.
Description
:
Query syntax:MEASure:FILE:LOAD?
Return dataMICROTEST
<Command>
: MEASure:BIAS:VOLTage<NR2>
:
MEASure:BIAS:VOLTage?
Set parameter: The value of BIAS Voltage -12.000~12.000
Function
:
It
sets or queries the BAIS Voltage.
Description
:
Set syntax:MEASure:BIAS:VOLTage -12
Query syntax: MEASure:BIAS:VOLTage?
Return data -1.01
<Command>
:
MEASure:BIAS:STATe<ON|OFF|0|1>
:
MEASure:BIAS:STATe?
Function
:
It
sets or queries the BIAS Voltage enabled/disabled.
Description
:
Set syntax:: MEASure:BIAS:STATe 1
Query syntax: MEASure:BIAS:STATe?
Return data When the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
<Command>
:MEASure:FILE?
Function
:
It queries all of the meter mode's filenames that are stored in the memory.
Description
:
Query syntax:
MEASure:FILE?
Return data
N, filename1, filename2, filename3, ..., filenameN
where N= total file count
<Command>
:
MEASure:STATistic {ON|OFF|0|1}
:
MEASure:STATistic?
Function
:
It sets or queries the statistic function in the meter mode.
Description
:
Set parameter {ON|OFF|0|1}
Set syntax:
MEASure:STATistic
1
Query syntax:
MEASure:STATistic?
Return data
0|1
where 0 = statistic function is disabled.