6.2 Short Circuit Correction
In CORRECTION page, <OFF><ON> function key to enable the short circuit cor-
rection, and use <Short> function key to do the short circuit correction. Before
starting the short circuit correction, please insert the shorting plate into the fixture
for executing the short circuit correction. If the correction fails or the figure is in-
correct, the page will display the calibration failure message; if successful, it will
return to the CORRECTION page automatically.
Test Cable Short Circuit Method
Fixture Short Circuit Method (lock the reed tightly after inserting the shorting bar)
NOTE:
If short circuit correction fails, you should check the following
points before starting the correction process again:
⚫
Check that the test fixture or the test cables are properly
connected.
⚫
Check that the test fixture or the test cables are properly
shorted together with the shorting bar. (Short circuit cor-
rection cannot be performed while any test piece is con-
nected to the test fixture or the test cables.)
⚫
Check the test leads are arranged as disable as possible
to their configuration in which measurement will be per-
formed.
⚫
During the correction process, be sure not to disturb the
fixture and the test cables or to move your hand near
them.