53
<Command>
*ESR?
Function
:
It queries the status of Standard Event Status Register, clean the Register after ex-
ecuted.
Description
:
Query syntax*ESR?
Return data<NR1>
<Command>
*IDN?
Function
:
It queries the device basic data.
Description
:
Return dataMICROTEST,6630-30G,0,1.194<field1>,<field2>,<field3>,<field4>
<field1>manufacturer
<field2>model number
<field3>serial number or 0
<field4>firmware revision
MICROTEST,6630-30G,0,1.194
<Command>
*OPC
Function
:
It sets Standard Event Status Register to 0 bit,when all operations are completed.
<Command>
*OPC?
Function
:
When all operations are completed, 1 will be returned.
Query syntax*OPC?
Return data 1
<Command>
*OPT?
Function
:
It queries the installed in the system interface.
Description
:
Query syntax*OPT?
If 6630 MAX test frequency is 1MHz , it returnsF01
If 6630 MAX test frequency is 3MHz, it returns F03
If 6630 MAX test frequency is 5MHz, it returns F05
If 6630 MAX test frequency is 10MHz, it returns F10
If 6630 MAX test frequency is 15MHz, it returns F15
If 6630 MAX test frequency is 20MHz, it returns F20
If 6630 MAX test frequency is 30MHz, it returns F30
<Command>
*RST
Function
:
It aborts all pending operations, and sets 6630 to its initial setups.
Description
:
Initial Setups
Parameters:Ls,Q,Z,
θdeg
FREQuency:1kHz
LEVEL:1Vac
SPEED:MED.
TRIGger:REPEAT