67
<Command>
:CORRection:LOAD:SPOT <spot number 1-16>
:CORRection:LOAD:SPOT?
Function
:
It sets or queries the load correction spot number that has been currently edited.
Description
:
Set parameter<spot number 1-16> (format is in NR1)
Set syntax:
CORRection:LOAD:SPOT 1
Query syntax:
:CORRection:LOAD:SPOT?
Return data
1
<Command>
:CORRection:LOAD:FREQuency <frequency in NR3>
:CORRection:LOAD:FREQuency?
Function
:
It sets or queries the load frequency for the current load correction spot.
Description
:
Set parameter<frequency> (format is in NR3),
where 0= FREQUENCY is OFF.
10HZ~30MHZ = FREQUENCY is ON.
Set syntax:
CORRection:LOAD:FREQuency 100k
Query syntax:
CORRection:LOAD:FREQuency?
Return data
+1.05 (Format is in NR3).
<Command>
:CORRection:LOAD:FUNCtion {LS-Q|LS-RS|LP-Q|LP-RP|CS-D|CS-RS|CP-
D|CP-RP|R-X|G-B|Z-DEG|Y-DEG}
:CORRection:LOAD:FUNCtion?
Function
It sets or queries the load function for the current load correction spot.
Description
:
Set parameter{LS-Q|LS-RS|LP-Q|LP-RP|CS-D|CS-RS|CP-D|CP-RP|R-X|G-B|Z-DEG|Y-DEG}
Set syntax:
CORRection:LOAD:FUNCtion LS-Q
Query syntax: CORRection:SPOT:FUNCtion?
Return data
LS-Q
<Command>
:CORRection:LOAD:REFerence <reference A,reference B>
:CORRection:LOAD:REFerence?
Function
It sets or queries the load reference values for the current load correction spot.
Description
:
Set parameter reference A,reference B (format is in NR3)
Set syntax:
CORRection:LOAD:REFerence 2E-03,40
Query syntax: CORRection:SPOT:FUNCtion?
Return data
+2.000000E-03,+4.01
<Command>
:CORRection:LOAD:VALue <load A,load B>
:CORRection:LOAD:VALue?
Function
It sets or queries the load values for the current load correction spot.
Description
:
Set parameter load A,load B (format is in NR3)
Set syntax:
CORRection:LOAD:VALue 2E+03,0.4
Query syntax: CORRection:LOAD:VALue?
Return data
+2.03,+4.000000E-01