Read the measured value
Read the measured value
Read Data format under Meter Mode
<para 1 data>,<para 2 data>,<para 3 data>,<para 4 data>,<status>,<bin number>,<para
1 compare status>,
<para compare status>,<para 3 compare status>,<para 4 compare status>
para 1-4 data
During value measuring, not all of four values will be displayed. The value will be dis-
played when the enableing display parameters are available. For example, if only two
parameters are enableed, then only two count of value will be transmitted.
status
– Measuring status, and the weighted value of each status refers to the final
value.
0
– Normal status without special status and without comparison.
1
– Measuring schedule error
2
– ALC error
4
– Other errors
8
– Reserve
16
– All parameters OK
32
– Some parameters NG
bin number
– Categorization result, and such value will not be displayed when closing
the bin function.
-1
– bin out, not in the categorization number
1
– 9 – bin number, the categorization result is 1-9.
para compare status 1-4
Measuring comparison result: If any parameter comparison function is enableed, then
the comparison result will be displayed for all of the enableed parameters.
0
– No comparison
1
– Parameter comparison result OK
2
– Parameter comparison result NG
Value transmit mode under Multi-step Test Mode.
<result>,<direction>,<bin number>,<step 1 result>,<step 1 data>,<step 2 result>,<step 2
data>,
<step 3 result>,<step 3 data>........<step n result>,<step n data>
result
– Finally judged test result
0
– Test not interrupted. Not completed
1
– Test steps all OK
2
– NG occurs to the test step
bin number - categorization result, and such value will not be displayed when closing the
bin function.
-1 - bin out not in the categorization number
1-9 - bin number the categorization result is 1-9
step result
0
– no test
1
– test OK
2
– test NG
step data
test value