69
<Command>
:LIST:CURRent <current NR3/disc >
:LIST:CURRent?
Function
:
It sets or queries the measurement current of list mode.
Description
:
Set parameter
When RO(output impedance) is setting 100Ω, the setup range of AC current
is {0.0001
~0.02|MAXimum|MINimum}. When RO(output impedance) is setting 25Ω, the
setup range of AC current is {0.0004~0.04|MAXimum|MINimum}. The range of DC current is
{0.0004~0.04|MAXimum|MINimum}.
Set syntax:LIST:CURRent 0.01
:LIST:CURRent 10M
:LIST:CURRent 10E-3
:LIST:CURRent MAXimum
:LIST:CURRent MINimum
Query syntax:LIST:CURRent?
Return data2.000000E-04 (Format is in <NR3>)
When the Level setup is current mode, the data returns 9.9E37.
<Command>
:LIST:SPEEd {MAXimum|FAST|MEDium|SLOW|SLOW2|0|1|2|3|4}
:LIST:SPEEd?
Function
:
It sets or queries the SPEED of list mode.
Description
:
Set parameterMAXimum/0, FAST/1, MEDium/2, SLOW/, SLOW2/4
Set syntax:LIST:SPEEd 1
:LIST:SPEEd FAST
Query syntax:LIST:SPEEd?
Return data FAST (Format is in <disc>)
<Command>
:LIST:DELay <delay time NR3/disc>
:LIST:DELay?
Function
:
It sets or queries the DELAY of list mode.
Description
:
Set parameterThe range of delay timeis {0.000~5.000|MAXimum|MINimum}.
Set syntax:LIST:DELay 0.5
:LIST:DELay 500m
:LIST:DELay 5E-3
:LIST:DELay MAXimum
:LIST:DELay MINimum
Query syntax:LIST:DELay?
Return data 0.000 (Format is in <NR2>)
<Command>
:LIST:COMParator:MODE <ABSolute|DEViation|PERCent|0|1|2>
:LIST:COMParator:MODE?
Function
:
It sets or queries the COMP MODE of list mode.
Description
:
Set parameterABSolute,0|DEViation,1|PERCent,2
Set syntax:MEASure:COMParator:MODE PERCent
Query syntax:LIST:COMParator:MODE?
Return data PERC (Format is in <disc>)
When the setup is 0, it returns ABS , When the setup is 1, it returns DEV , When the setup is 2,
it returns PERC.