18
•
The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance with
the operating instructions.
Please contact your local Microlife-Service (see foreword).
10. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Operating conditions:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
130 g (including batteries)
Dimensions:
80 x 70 x 70 mm
Cuff size:
13.5 - 21.5 cm (5.25 - 8.5 inches)
Measuring procedure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
2 x 1.5 V alkaline batteries; size AAA
Battery lifetime:
approx. 460 measurements
(using new batteries)
IP Class:
IP20
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life:
Device: 5 years or 10000 measurements
Accessories: 2 years