Page 54
Operation
MSC7401 / 7802 / 7602 Miniature sensor controller for inductive displacement sensors
5.4.2
Zero-Point Search
D1:
Channel
D2:
E1
D3:
E2
D4:
Value
G
O
LED off
Set the output signal U
0
.
6 VDC or 12 mA is preset.
ENTER
G
R
G
O
The LED flashes
and color changes
depending on the
output signal
(green = too low
red = too high).
O
Move the measuring object to position X
0
until the
output has reached U
0
.
Lights orange when
the measuring object
is in the electrical
center of the sensor.
ENTER
G
O
The LED color
changes depending
on the position
of the measuring
object.
O
Move the measuring object to position X
2
1
and
change the output signal U
2
with
Flashes orange when
the measuring object
is in the electrical
center of the sensor.
Fig. 55 Menu structure for the MSC 7401 / 7802 / 7602 controllers, adjustment mode: Zero-point search
1) Position X
2
must
be > 10 % of the
measuring range
away from X
1
.