Page 85
Operation
colorSENSOR CFO
5.8
Factory Reset
A factory reset can be accomplished as follows:
Disconnect the sensor from the power supply.
Hold the
INT
and
ESC
keys depressed.
Avoid pressing any other keys during this procedure. Otherwise the process can become stuck in an unde-
fined state.
> Damage to sensor
Switch the power supply back on.
Release the keys.
The sensor then starts to reset to the factory default settings. During this operation all sensor settings as well
as all firmware updates performed in the meantime are discarded. Resetting requires approx. 5 minutes.
In the event of an error, e.g. interruption in power supply, the reset is continued during the next restart. No
sensor states are signaled and no entries are possible during the reset operation. Pressing a key is acknowl-
edged by error flashing (3x100 / 300 ms).
Indicators
Output LEDs:
Pulsating progress bar (indeterminate progress bar)
Key LEDs:
Not illuminated
Power-LED:
Illuminated continuously
Key lock LED:
Illuminated continuously
NOTICE