4 Start-up
947 Professional UV/VIS Detector Vario (2.947.00x0)
■■■■■■■■
37
Figure 27
Optimized intensity spectrum with UV lamp
6
Click on
[OK]
.
The lamp values are transferred to the instrument. The window is
closed. The adjustment is complete.
4.1.3
Operation with VIS lamp
NOTICE
Only follow the following steps if the instrument is in operation exclu-
sively with the VIS lamp.
The lamp settings must be adjusted if the VIS lamp was replaced or if
checking the intensity spectrum after first start-up displays a cut-off
Remove the UV lamp completely
. Close the
opening of the lamp with the socket cap. Otherwise the lamp is automati-
cally switched on as soon as the detector settings are opened.
Adjusting the lamp settings
Prerequisites
■
The instrument is operated exclusively with the VIS lamp.