3.5 Installing the lamps
■■■■■■■■■■■■■■■■■■■■■■
24
■■■■■■■■
947 Professional UV/VIS Detector Vario (2.947.00x0)
3
Lamp socket
4
Groove for the alignment of the lamp
5
Lamp surface
Inserting the UV lamp
Prerequisite
■
The instrument is switched off.
Accessories
■
UV lamp (6.2804.110)
The UV lamp is not part of the products 2.947.0010 and 2.947.0020.
The UV lamp must be purchased separately.
1 Removing the socket cap for the UV lamp
Loosen the adjustment ring
Remove the adjustment ring and the socket cap.
Figure 17
Removing the socket cap for the UV lamp
2 Inserting the UV lamp
Hold the UV lamp on the lamp socket
on the lamp cooling element.
Align the groove on the lamp socket
with the positioning pin
in the lamp cooling element. The positioning pin is located on that
side in the lamp cooling element that is closer to the rear of the
instrument.
Push the UV lamp into the opening for the UV lamp until it stops.
Summary of Contents for 2.947.0010