6 Operation and maintenance
947 Professional UV/VIS Detector Vario (2.947.00x0)
■■■■■■■■
53
If the noise on the baseline persists, the flow-through cell must be
disassembled and cleaned manually
(see "Dismantling and cleaning
the flow-through cell manually", page 53)
Dismantling and cleaning the flow-through cell manually
NOTICE
Ex works, the retaining screws are tightened with a torque wrench. This
guarantees a pressure stability up to 5 MPa (50 bar).
Once the flow-through cell has been opened and closed again man-
ually, this maximum pressure stability is no longer guaranteed.
Prerequisites
■
Rinsing the flow-through cell with solvent was ineffective.
Accessories
■
Slotted screwdriver, size 5
■
Seals (6.2764.000)
1 Removing the flow-through cell
■
Disconnect the inlet capillary and the outlet capillary.
■
Loosen and remove the knurled screws.
■
Remove the flow-through cell from the optical block.