4 Start-up
947 Professional UV/VIS Detector Vario (2.947.00x0)
■■■■■■■■
39
Figure 28
Optimized intensity spectrum with VIS lamp
4
Check the new intensity spectrum that is displayed.
Click on
[OK]
.
The adjusted lamp values are transferred to the instrument. The win-
dow is closed. The lamp optimization is complete.
4.2
Checking the intensity spectrum
Check the intensity spectrum in the following cases:
■
After initial start-up
■
After replacing a lamp
■
If increased noise or measurement inaccuracies occur after prolonged
use
Checking the intensity spectrum
Prerequisite