3.5 Installing the lamps
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
947 Professional UV/VIS Detector Vario (2.947.00x0)
Slide the pressure screw over the detector outlet capillary such that a
small part of the capillary is visible at the top.
Tighten the capillary in the detector outlet using the pressure screw.
NOTICE
The detector outlet capillary (6.1831.100) is 1 m long and may not
be shortened.
3.5
Installing the lamps
CAUTION
Damage to the lamp
Reduced light transmission due to residues on the lamp surface. Dam-
age to the lamp due to burnt-on residues on the lamp surface.
■
Do not touch the lamp surface with your fingers!
■
If the lamp surface is contaminated, clean the lamp with alcohol and
a lint-free cloth.
■
Make sure the lamp surface is dry before using the instrument.