6.5 Cleaning the flow-through cell
■■■■■■■■■■■■■■■■■■■■■■
54
■■■■■■■■
947 Professional UV/VIS Detector Vario (2.947.00x0)
Figure 39
Flow-through cell – parts
1
Retaining screw
2
Outer seal
The outer seal is attached to the retaining
screw. When removing the flow-through
cell, the outer seal does not have to be
removed from the retaining screw.
3
Lens
4
Inner seal
5
Measuring cell – opening
6
Cell holder
2 Removing the lens
Loosen the retaining screw
with a slotted screwdriver, size 5,
and remove it.
The outer seal is attached to the inside of the regaining screw
Remove the outer seal.
Carefully remove the lens
from the
measuring cell.
Put the lens onto a white paper. This allows you to see possible
deposits. There are deposits that only absorb in the UV range. These
deposits are not visible.
3 Cleaning the lens
Clean the lens with ultrapure water, methanol or another suitable
solvent (depending on the application) and wipe with a lint-free
cloth.