4 Operation
792 Basic IC
70
Ident
User defined identifier (title) for the chroma-
togram to be displayed in the title bar of the
chromatogram window and in the
Chromatogram
open
window.
Calibration level
Calibration level (0 = sample; 1
}
n = calibra-
tion solutions).
Info 1 / Info 2
Sample
description.
Volume
Injected volume in
P
L.
Dilution
Dilution of the sample.
Vial number
Autosampler vial position to take sample from.
Amount
Sample amount. If this value is different for the
calibration run (
c
) and the sample run (
s
), the
component concentrations of the sample are
calculated as follows:
C
s
=
C
c
•
Amount
s
/
Amount
c
Internal standard amount
Concentration of the internal standard compo-
nent for relative concentration calculations.
Date/time when...
Date and time of sample collection (the default
values are equal to the date and time when the
chromatogram starts).
SYSTEM / Control / Stop determination
Stop running determination
. Data acquisition and time program are
terminated immediately. The recorded chromatogram is saved auto-
matically if the
Save chromatogram after the run
option on the
Passport /
Processing
tab is enabled.
Alternatively the determination can be stopped by clicking the icon
of the chromatogram window. In this case, the user is asked always if
the determination should be saved or not.
Print system parameters
SYSTEM / System / Print parameter
A report of the system parameters including the time program is cre-
ated and opened using the
«
Microsoft Word
»
program. The
*.rtf
file
opened can be printed, saved and exported into other programs. The
system report includes the name of the method linked to the system,
the configuration settings for the 792 Basic IC, the system startup val-
ues and the time program (incl.
Remote configuration
), if such a program
exists and is switched on (
ENABLED
).