2.5 Connection to the PC
792 Basic IC
21
x
Click on
COM1
using the right mouse button and select the
Change
menu item to open a window listing all available COM
ports at the PC.
x
Select the desired COM port to which the 792 Basic IC has
been connected and click on
<OK>
. The window is closed.
x
Close the
Links
window by clicking on
<OK>
.
3 Open a system
x
Click on
File / Open / System
in the main window. Open the
folder
Default
. Select the system file
Default.smt
and click on
<Open>
. The corresponding system window is opened:
x
If the connection between PC and 792 Basic IC is working the
message
Hardware settings file for 792 unit with serial number
'#####' not found! Create?
will appear. Click
<Yes>
in order to
create the configuration file
#####.792
for this instrument.
x
If the connection between PC and Personal IC does not work
then the message
Detection of hardware failed[792 Basic IC [
COM# ]]
appears in the
SYSTEM STATE
window. In this case
check whether the instrument has been switched on, whether
the connection cable is connected up properly and whether
the COM interface has been set correctly (see point
2
). Then
repeat point
3
.