5 Notes – Maintenance – Faults
792 Basic IC
152
2 Dismantle suppressor
x
Unscrew nut
102
102
102
102
from suppressor holder
105
105
105
105
.
x
Pull out connection piece
103
103
103
103
and suppressor rotor
104
104
104
104
from
suppressor holder
105
105
105
105
(the connection piece and the rotor
normally stick together).
x
Loosen connection piece
103
103
103
103
from suppressor rotor
104
104
104
104
.
3 Clean input and output leads
x
Connect each of the 6 capillary tubings attached to connec-
tion piece
103
103
103
103
to the high-pressure pump one after another,
and pump through ultrapure water.
x
Check whether solution emerges from connection piece
103
103
103
103
.
If one of the input or output leads remains blocked, replace
the connection piece
103
103
103
103
(order number 6.2832.010).
4 Clean suppressor rotor
x
Clean the sealing surface of suppressor rotor
104
104
104
104
using a
lint-free cloth and ethanol.
5 Insert suppressor rotor
x
Insert suppressor rotor
104
104
104
104
in suppressor holder
105
105
105
105
in such
a way that the tubing connections at the rear of the rotor fit in
the corresponding openings inside the rotor, and that one of
the three holes in the rotor can be seen from below in one of
the openings of the holder.
x
If the rotor has been inserted correctly, its sealing surface will
be about 4 mm inside the holder. If this is not the case, bring
the rotor into the correct position from below with the aid of a
sharp object (e.g. a screwdriver).
6 Clean connection piece
x
Clean the sealing surface of connection piece
103
103
103
103
using a
lint-free cloth and ethanol.
7 Insert connection piece
x
Insert connection piece
103
103
103
103
in suppressor holder
105
105
105
105
in such
a way that connection "1" is at the top, and that the three lugs
on the connection piece fit in the corresponding openings of
the holder.
x
Screw nut
102
102
102
102
onto the thread of suppressor holder
105
105
105
105
manually (do not use tools).
8 Connect and condition the suppressor
x
Reconnect the suppressor to the IC system.
x
Before switching the suppressor to the next position for the
first time, rinse all 3 suppressor units with solution for 5 min.