4.4 Methods
792 Basic IC
99
Identification
Identification
parameters.
Reference peaks
Identification parameter for reference compo-
nents. Default is
Height
.
Other peaks
Identification parameter for ordinary compo-
nents. Default is
Time
.
Retention units
Choice of retention units. Default is
min
.
Retention time
Retention
time.
<Update>
Updates expected retention times of compo-
nents according to the current chromatogram.
Worst case...
Information on the component with the worst
(largest) deviation of actual and expected reten-
tion time. Deviation is given as a part of com-
ponent’s identification window.
Average relative deviation
Relative deviation averaged for all components.
Summary of Contents for 2.792.0020 Basic IC
Page 53: ...2 Installation 792 Basic IC 44...
Page 197: ...6 Appendix 792 Basic IC 188...